DESCRIPTION
The leaded wall pass-through can be used to temporarily store under protection injectable doses from the preparation room for collection from the injection room.
In order to secure the transfer of radiopharmaceutical doses from the preparation laboratory or the radiopharmacy to the administration rooms, Lemer Pax has developed the double door shielded wall-pass-through type GTD02, designed with a 2 mm lead shield and with standard dimensions L 625 x D 600 x H 750 mm.
Complete safety is ensured by the option of mechanical or electrical interlocking of the doors which can be combined with the option of an “overpressure pass-through” equipped with a HEPA filter thus preventing any unwanted transfer of particles from one room to another when using the pass-through.
The “oculus” option offers the user permanent visual control of patient sources or doses ready for use or administration, through a lead glass window integrated into the pass-through doors.
The pass-through is modular, and other dimensions and shielding are also possible based on a feasibility study carried out by the Lemer Pax R&D team to adapt to all isotopes and their packaging and thus provide the radiation protection adapted to your applications.
CHARACTERISTICS
ADDITIONNAL PRODUCT