WHY CHOOSE TMP shielded hot cell

  • GMP conformity: meets the good manufacturing practice standards
  • Filtration: air quality in accordance with ISO standard 14644-1, HEPA filters (inlet) / HEPA + active charcoal (outlet)
  • Sealing: sealing Class 3 in accordance with ISO standard 10648
  • Appropriate radiation protection: between 100 and 200 lead shielding thickness according to your needs
  • Manipulation: equipped with two remote manipulators and/or two handling tongs with ball joints, the cell is configured to perform the necessary tasks or operations
DESCRIPTION
CHARACTERISTICS
ADDITIONNAL PRODUCT

DESCRIPTION

Designed to enable operators to carry out a maximum number of hot operations, the TMP shielded hot cell can be customized to suit your needs and applications (research and production).

A highly shielded cell for all of your applications

This highly shielded cell is equipped with remote manipulators for more complex manipulations. It can also be fitted with remote handling tongs for simpler tasks.

The TMP shielded hot cell is fully compliant with regulatory requirements. Both in terms of Good Manufacturing Practice (GMP) and radiation protection standards.

Their meticulous interior finish (no asperity Ra < 0.8 μm) in 316L stainless steel and their radiused corners are ideal for cleaning and decontamination.

Connection to the building’s nuclear ventilation network is ensured by automatically opening valves. Thus enabling the hot cell to be put under negative pressure or confined. A pressure gauge continuously displays the vacuum level in the enclosure*. Filtration is provided by a HEPA filter at the inlet. At the outlet, by one or more HEPA and activated carbon filters. The latter are installed inside the hot cell as standard.

All the functions of the cell are controlled on a touch screen: inflating of the seals, lighting, ventilation valve control, etc. The screen also continuously displays the irradiation values and negative pressure inside each hot cell*.

TMP shielded hot cell allows to carry out a maximum number of hot operations

The front of the hot cell is equipped with a large leaded-glass viewing window (dimensions: W 460 x H 460 mm) for full visibility of the work area. The interior of the hot cell (with a work surface width of between 1,000 and 2,000 mm, and a depth of between 800 and 1,200 mm) offers a comfortable, fully accessible workspace, which can be accessed using handling tongs with ball joints and/or suitably installed remote manipulators.

Access to the work area is via a large rear door. It is located in the technical gallery, for all loading and/or cleaning operations. This configuration separates the working area (on the laboratory side) from the rear area (on the technical side), with different pressure and dust levels.

Each hot cell features liquid or gaseous target tube inlets, and different types of electric and fluidic connectors (nitrogen, hydrogen, compressed air, etc.), and covers all users needs.

They are also equipped with individually controllable electrical outlets.

The cell can be equipped with a large armored airlock or a simple automatic sliding door inside the cell to allow products to enter or leave. The airlock features an automatically operated loading drawer to facilitate transfer operations.

Depending on the application, this cell can also be fitted with :

  • DPTE® airlock and PMP doors
  • garbage can compartment for liquid or solid waste
  • dose calibrator location
  • bottle outlet

CHARACTERISTICS

  • Overall dimensions: depend on the inner chamber dimensions
  • Hot cell weight: related to the shielding thickness and the inner chamber dimensions
  • Exterior finish: white lacquered aluminium finish, similar to cleanroom panels, and 304L stainless steel
  • Connection to the nuclear ventilation network: DN50 per chamber – approx. 50 m3/h
  • General power supply voltage: 230 V / 32 A - 50 / 60 Hz
  • Exchange of signals: several dry contacts are provided for communication with the system. 4-20mA signals can be supplied.
  • Compressed air: 7 bars - Qualité d'air 2-4-2 suivant ISO 8573-1
  • Technical gas: according to requirements
  • Floor load: related to cell weight